Technical Specifications
CITREX H3
Flyer Technical Specifications
Flow and pressure measurements | ||
Flow | High Flow |
±300 L/min*** ±1.9 %* or ±0.1 L/min** |
Measuring bidirectional | yes | |
Temperature compensated | automatic | |
Pressure compensated | automatic | |
Humidity compensated | manually | |
Pressure | ||
Flow channel | PChannel |
-50–150 mbar ±0.75 %* or ±0.1 mbar** |
Atmospheric pressure | PAtmo |
500–1150 mbar ±1 %* or ±5 mbar** |
Units | ||
Flow | L/min, L/s, cfm | |
Pressure | inH2O, bar, mbar, cmH2O, mmHg | |
Other measurements | ||
Oxygen (Option) (pressure comp. ≤ 150 mbar) |
O2 |
0–100 % ±1 % O2** |
Gas temperature | Temp. |
0–50°C ±1.75 %* or ±0.5 °C** |
Gas types | Air, Air/O2, O2, N2O, N2O/O2 | |
Gas standards | ATP, ATPD, ATPS, AP21, STP, STPH, BTPS, BTPS-A, BTPD | |
Ventilation parameter | ||
Breath rate | Rate |
1–1000 bpm ±1 bpm* or ±2.5 %** |
Time | Ti |
0.05–60 s ±0.02 s |
Ratio | I:E |
1:300–300:1 ±2.5 %* |
Breath volume | V | ±2 %* or ±0.20 mL (>6 L/min)** |
Tidal Volume | Vti |
±10 L ±2 %* or ±0.20 mL (>6 L/min)** |
Minute Volume | Vi |
0–300 L/min ±2.5 %* |
Peak flow | PFInsp, PFExp |
± 300 L/min ±2.0 %* or ±0.1 L/min** |
Pressure | PPeak, PMean, PEEP |
0–150 mbar ±0.75 %* or ±0.1 mbar** |
General information | ||
Display | 1.7'' with touch control elements (color), Realtime curves | |
Data storage | microSD Card | |
Power | 100–240 VAC, 50–60 Hz | |
Dimension (w x d x h) | 11.4 x 7 x 6 cm | |
Weight | 0.38 kg | |
Battery | 4 hours | |
Approvals | CE, BC (Energy Efficiency for Battery Charging Systems), CSA (North America), IEC 61010-1:2010, IEC 61326-2:2012 | |
The greater tolerance is valid: |