Technical Specifications


CITREX H3


Flyer Technical Specifications

Flow and pressure measurements
Flow High Flow ±300 sL/min***
±1.9 %* or ±0.1 sL/min**
Measuring bidirectional   yes
Temperature compensated   automatic
Pressure compensated   automatic
Humidity compensated   manually
Pressure
Flow channel PChannel -50–150 mbar
±0.75 %* or ±0.1 mbar**
Atmospheric pressure PAtmo 500–1150 mbar
±1 %* or ±5 mbar**
Units
Flow   L/min, L/s, cfm
Pressure   inH2O, bar, mbar, cmH2O, mmHg
Other measurements
Oxygen (Option)
(pressure comp. ≤ 150 mbar)
O2 0–100 %
±1 % O2**
Gas temperature Temp. 0–50°C
±1.75 %* or ±0.5 °C**
Gas types   Air, Air/O2, O2, N2O, N2O/O2
Gas standards   ATP, ATPD, ATPS, AP21, STP, STPH, BTPS, BTPS-A, BTPD
Ventilation parameter
Breath rate Rate 1–1000 bpm
±1 bpm* or ±2.5 %**
Time Ti 0.05–60 s
±0.02 s
Ratio I:E 1:300–300:1
±2.5 %*
Breath volume V ±2 %* or ±0.20 mL (>6 sL/min)**
Tidal Volume Vti ±10 L
±2 %* or ±0.20 mL (>6 sL/min)**
Minute Volume Vi 0–300 sL/min
±2.5 %*
Peak flow PFInsp, PFExp ± 300 sL/min
±2.0 %* or ±0.1 sL/min**
Pressure PPeak, PMean, PEEP 0–150 mbar
±0.75 %* or ±0.1 mbar**
General information
Display   1.7'' with touch control elements (color), Realtime curves
Data storage   microSD Card
Power   100–240 VAC, 50–60 Hz
Dimension (w x d x h)   11.4 x 7 x 6 cm
Weight   0.38 kg
Battery   4 hours
Approvals   CE, BC (Energy Efficiency for Battery Charging Systems), CSA (North America), IEC 61010-1:2010, IEC 61326-2:2012

The greater tolerance is valid:
* Tolerance related to the measured value,
** Absolute tolerance,
*** The unit sL/min is based on ambient conditions of 0 °C and 1013.25 mbar (DIN 1343).

You want to benefit from the advantages of the CITREX H3 every day?

Get quotation