Technical Specifications
CITREX H5
Flyer Technical Specifications
Flow and pressure measurements | ||
Flow | High Flow |
±300 L/min ±1.9 %* or ±0.1 L/min** |
Measuring bidirectional | yes | |
Temperature compensated | automatic | |
Pressure compensated | automatic | |
Humidity compensated | manually | |
Pressure | ||
High / Vaccum | PHigh |
-1−10 bar ±1 %* or ±7 mbar** |
Differential | PDiff |
±200 mbar ±0.75 %* or ±0.1 mbar** |
Flow channel | PChannel |
-50−150 mbar ±0.75 %* or ±0.1 mbar** |
Atmospheric pressure | PAtmo |
500−1150 mbar ±1 %* or ±5 mbar** |
Units | ||
Flow | L/min, L/s, cfm, mL/min, mL/s | |
Pressure | bar, mbar, cmH2O, inH2O, Torr, inHg, hPa, kPa, mmHg, PSI | |
Other measurements | ||
Oxygen (pressure comp. ≤ 150 mbar) |
O2 |
0−100 % ±1 % O2** |
Gas temperature | Temp. |
0−50°C ±1.75 %* or ±0.5 °C** |
Gas types | Air, Air/O2, O2, N2O, N2O/O2, CO2, N2, He/O2 | |
Gas standards | ATP, ATPD, ATPS, AP21, STP, STPH, BTPS, BTPS-A, BTPD, BTPD-A, 0/1013, 20/981, 15/1013, 25/991, 20/1013, NTPD, NTPS | |
Ventilation parameter | ||
Breath rate | Rate |
1−1000 bpm ±1 bpm* or ±2.5 %** |
Time | Ti, Te |
0.05−60 s ±0.02 s |
Ratio | I:E |
1:300−300:1 ±2.5 %* |
Ti/Tcyc |
0−100 % ±5 %* |
|
Breath volume | V | ±2 %* or ±0.20 mL (>6 L/min)** |
Tidal volume | Vti, Vte |
±10 L ±2 %* or ±0.20 mL (>6 L/min)** |
Minute volume | Vi, Ve |
0−300 L/min ±2.5 %* |
Peak flow | PFInsp, PFExp |
± 300 L/min ±1.9 %* or ±0.1 L/min** |
Pressure | PPeak, PMean, PEEP, PPlateau, IPAP |
0−150 mbar ±0.75 %* or ±0.1 mbar** |
Compliance | CStat |
0−1000 mL/mbar ±3 %* or ±1 mL/mbar** |
General information | ||
Realtime curves | yes | |
Display | 4.3'' Multi-Touch (color) | |
Interface | RS-232, USB, Ethernet, CAN, Analog Out, TTL, WLAN | |
Data storage | Internal and microSD Card | |
Power | 100−240 VAC, 50–60 Hz | |
Dimension (w x d x h) | 11.4 x 7 x 7,3 cm | |
Weight | 0.52 kg | |
Battery | 5 hours | |
Approvals | CE, BC (Energy Efficiency for Battery Charging Systems), CSA (North America), IEC 61010-1:2010, IEC 61326- 2:2012 | |
The greater tolerance is valid: * Tolerance related to the measured value, ** Absolute tolerance, *** The unit L/min is based on ambient conditions of 0 °C and 1013.25 mbar (DIN 1343). |